Specifications
- Temperature range Ambient +5°C to 375°C
- Resolution 0.1 °C/°F
- Accuracy Over full range better than +/- 0.4°C, typ. +/- 0.2°C
- Stability +/- 0.05°C
- Heating time Ambient to 300 °C in 5.75 minutes
- Cooling time 300°C to 60°C in 14 minutes
- Stabilization 5 minutes
- Well dimensions Depth 100 mm, Diameter 13 mm
- Switch test Powerless switch, LED indicator, 5 VDC, 16 mA max
- Communication RS232 for use with e.g. AutoCal
- Supply voltage 220Volt, 50/60 Hz or 110 Volt, 50/60 Hz. Selectable
- Power consumption 250 Watts
- Fault protection Internal sensor break detection
- Fuse 3.15 A. (slow)
- Thermal Protection switch 75 °C
- Temp switch test 5 VDC, 16 mA max.
- Safety NEN –EN-IEC 601010
- Conformity (EC) to NEN-EN-IEC 61000-4-3, -3-2, -3-3, -4-2, -4-4, 4-5
- Weight approx. 1740 grams, without power cord, incl. 6.2 mm insert
- Dimensions w 149 x d 159 x h 74 mm.
- Standard scope of supply Traqc-3FM, 1 insert with 6mm bore, standard calibration Certificate, Powerlead
- Options Traqc-3 robust rugged blue carrying case, additional inserts with bore: 3.5, 6.5, 4,5, undrilled. Others on request. Accredited calibration certificate traceable to international standards
|
|