Tradinco Traqc-20 LPC Low Pressure Calibrator

Tradinco Traqc-20 LPC Low Pressure Calibrator

Product Description

 
 
 
   
 
The LPC measures and generates pressures and consequently it is more than a mere measuring instrument – it can also be applied as test and calibration instrument for pressure sensors and pressure switches. Thanks to internal pressure generation no auxiliary tools are required for operation. An internal Li-ion battery is available for mobile operation. Parallel operation of mains and battery supply allows for maximum flexibility.
 
The Traqc-20 LPC is particularly suitable for stable pressure control in the very low pressure ranges (<<1000mbar), which is often used when environmental sensors need to be calibrated for cleanrooms, hospital areas and high-quality filters.
 
  • Very low pressure calibrator 0...1 to 0...1000 mBar
  • High precision
  • Compact size
  • Easily operated
  • Li-ion battery
 
 
Specifications
  • Pressure range 1 mbar, 10 mbar, 100 mbar, 1000 mbar
  • Accuracy 0.3 % FS, 0.1 % FS, 0.1 % FS, 0.1 % FS
  • Type of pressure Positive or negative gauge pressure or differential pressure
  • Pressure units Pa, kPa, hPa, bar, mbar, psi, inH2O, inHg, mmHg, Torr, mmH2O
  • Adjustment accuracy 0.05 %
  • Long-term stability < +/- 0.1% FS/year
  • Temperature drift ≤ 0.03 % FS/K 0.0 % with automatic of manual zero-point adjustment
  • Usable pressure range -10 - 110 %
  • Zero balance Automatic: in adjustable time intervals Manual: ZERO button
  • Instruments version desktop instrument with carry handle
  • Dimensions 102.6 x 257 x 271 mm without carry handle
  • Weight 4.5 kg
  • Ingress protection IP20
  • Pressure generation internal, electric pump
  • Screen Graphic display
  • Resolution 5 digits
  • Keyboard membrane keypad Menu
  • language German, English, Spanish and French
  • Connections Pressure connections 6.6 x 11 (hose diameter, D = 6 mm)
  • Power supply DC 24 V, 1 A
  • Battery type Li-Ion
  • Battery life approx. 8 h
  • Media Ambient air
  • Operating temperature 10 ... 40 °C
  • Storage temperature -10 ... 70 °C Specifications Di mensions
 
 
 
 
 
 
Copyright 2024 by Rex Instruments 201401004426 (1080500-W)